In situ sensing to enable the 2010 thermodynamic equation of seawater

Date

2017-01-03

Authors

Dakin, Del Thomas

Journal Title

Journal ISSN

Volume Title

Publisher

Abstract

The thermodynamic equation of seawater - 2010 (TEOS-10) is hampered by the inability to measure absolute salinity or density in situ. No new advances for in situ salinity or density measurement have taken place since the adoption of the practical salinity scale in 1978. In this thesis three possible technologies for in situ measurements are developed and assessed: phased conductivity, an in situ density sensor and sound speed sensors. Of these, only sound speed sensors showed the potential for an in situ TEOS-10 measurement solution. To be implemented, sensor response times need to be matched and the sound speed sensor accuracy must be improved. Sound speed sensor accuracy is primarily limited by the calibration reference, pure water. Test results indicate the TEOS-10 sound speed coefficients may also need to be improved. A calibration system to improve sound speed sensor accuracy and verify the TEOS-10 coefficients is discussed.

Description

Keywords

seawater density, water density, acoustic impedance, absolute salinity, sound speed, sound speed standard, sound speed calibration, phased conductivity, TEOS 10