Lead-doped scintillator dosimeters for detection of ultrahigh dose-rate x-rays

dc.contributor.authorHart, Alexander
dc.contributor.authorCecchi, Daniel
dc.contributor.authorGiguère, Cloé
dc.contributor.authorLarose, Frédérique
dc.contributor.authorTherriault-Proulx, François
dc.contributor.authorEsplen, Nolan
dc.contributor.authorBeaulieu, Luc
dc.contributor.authorBazalova-Carter, Magdalena
dc.date.accessioned2023-03-30T20:12:30Z
dc.date.available2023-03-30T20:12:30Z
dc.date.copyright2022en_US
dc.date.issued2022-05-11
dc.description.abstractObjective. Lead-doped scintillator dosimeters may be well suited for the dosimetry of FLASH-capable x-ray radiotherapy beams. Our study explores the dose rate dependence and temporal resolution of scintillators that makes them promising in the accurate detection of ultrahigh dose-rate (UHDR) x-rays. Approach. We investigated the response of scintillators with four material compositions to UHDR x-rays produced by a conventional x-ray tube. Scintillator output was measured using the HYPERSCINT-RP100 dosimetry research platform. Measurements were acquired at high frame rates (400 fps) which allowed for accurate dose measurements of sub-second radiation exposures from 1 to 100 ms. Dose-rate dependence was assessed by scaling tube current of the x-ray tube. Scintillator measurements were validated against Monte Carlo simulations of the probe geometries and UHDR x-ray system. Calibration factors converting dose-to-medium to dose-to-water were obtained from simulation data of plastic and lead-doped scintillator materials. Main Results. The results of this work suggest that lead-doped scintillators were dose-rate independent for UHDR x-rays from 1.1 to 40.1 Gy s−1 and capable of measuring conventional radiotherapy dose-rates (0.1 Gy s−1) at extended distance from the x-ray focal spot. Dose-to-water measured with a 5% lead-doped scintillator detector agreed with simulations within 0.6%. Significance. Lead-doped scintillators may be a valuable tool for the accurate real-time dosimetry of FLASH-capable UHDR x-ray beams.en_US
dc.description.reviewstatusRevieweden_US
dc.description.scholarlevelFacultyen_US
dc.description.sponsorshiphis work was partly supported by the Canada Foundation for Innovation, the British Columbia Knowledge Development Fund, NSERC Discovery Grants held by M B C and L B, an NSERC Discovery Accelerator Supplement Grant, NSERC USRA grants held by D C and C G, the New Frontiers in Research Fund, and the Canada Research Chair program.en_US
dc.identifier.citationHart, A., Cecchi, D., Giguère, C., Larose, F., Therriault-Proulx, F., Esplen, N., Beaulieu, L., & Bazalova-Carter, M. (2022). Lead-doped scintillator dosimeters for detection of ultrahigh dose-rate x-rays. Physics in Medicine & Biology, 67(10), 105007. https://doi.org/10.1088/1361-6560/ac69a5en_US
dc.identifier.urihttps://doi.org/10.1088/1361-6560/ac69a5
dc.identifier.urihttp://hdl.handle.net/1828/14924
dc.language.isoenen_US
dc.publisherPhysics in Medicine & Biologyen_US
dc.subject.departmentDepartment of Physics and Astronomy
dc.titleLead-doped scintillator dosimeters for detection of ultrahigh dose-rate x-raysen_US
dc.typeArticleen_US

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Hart_Phys._Med._Biol._2022.pdf
Size:
838.09 KB
Format:
Adobe Portable Document Format
Description:
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
2 KB
Format:
Item-specific license agreed upon to submission
Description: